Electrical Safety Tester- 10 Lead ECG
A comprehensive full function Electrical Safety Analyzer that tests the electrical safety of any medical equipment.
The LKG 610, Electrical Safety Tester is designed to test any piece of medical equipment. The full featured Electrical Safety Analyzer is microprocessor controlled and has a full compliment of ten leads.
The LKG 610, Electrical Safety Tester measures line voltage, outlet polarity, device current, earth leakage current, chassis resistance, and case leakage current. In addition it performs lead ISO tests and point to point measurements.
Tests are quickly and easily performed when the appropriate test function key is pushed. Measurements results are presented on the large LCD display and measurements units are clearly indicated by an LED. Large switches change the polarity, open and close neutral, and lift the ground. Ten universal lead connectors accept both snaps and any size posts.
Calibrated test points are available to ensure performance accuracy and repeatability. Either the AAMI and IEC 601 test loads may be selected. The receptacle is fused with an accessible 20 A fuse and the tester is fused internally.
The comprehensive LKG 610, Electrical Safety Tester provides the best combination of features, performance, and value of any portable electrical safety analyzer.
|
 |
 |
 |
Line Voltage |
 |
Cord Resistance |
 |
Case Leakage Current |
 |
Earth Leakage Current |
 |
Leads Leakage Current |
 |
Leads ISO Test |
 |
Instrument Current |
 |
Point-to-Point Measurement |

|
|
Measurement Range
Resistance: 0 - 1999 mW
Leakage Current: 0 - 1999 µA
Instrument Current: 0 - 19.99 A
Accuracy
Resistance |
: |
± 2% ± 1 LSD |
Leakage |
: |
± 1% of FS DC to 1 KHz ± 1 LSD,
± 2% of FS DC to 1 KHz to 100 KHz,
± 4% of FS 100 kHz to 1 MHz
|
Instrument Current |
: |
± 2% ± 1 LSD |
Controls
12 Push Button Function Keys
Open / Closed Neutral Switch
Normal / Off / Reverse Polarity Switch
Momentary Ground Open Switch
10 Universal Snap / Post ECG Connectors
Test Load Selector Switch
Calibrated Test Points
Reference Ground: 1W
Leakage: 200 µA
|
Display |
: |
± 2% ± 1 LSD |
Physical Dimensions |
: |
8.0 x 4.5 x 2.0 inches
(20.3 x 11.4 x 5.0 cm) |
Weight |
: |
2.5 lbs (1.1 kg) |
Power Requirements |
: |
110 VAC 50 - 60 Hz, 20A
220 VAC 50 - 60 Hz, 10 A
|
Environmental
Operating Range |
: |
59º to 95º F (15º to 35º C) |
Storage Temperature |
: |
0º to 122º F ( -18º to 50º C) |
Relative Humidity |
: |
90% (max) at temperatures |
|
|
 |
| Part Number |
Description |
Price |
535 |
LKG 610, 110 VAC |
|
535 - 220 |
LKG 610, 220 VAC |
|
535 - Case |
Soft Carrying Case |
 |
503 |
Test Cable |
 |
|
|
 |
|


 |